Series 901 Enhanced Ultrasonic Flow Meter Operations & Maintenance Manual REV 4/02

Seite 2 - Pipe Preparation

Rev. 04/02 -9- D901/M PART 2 - PRE-INSTALLATION CHECKOUT After unpacking, it is recommended to save the shipping carton and

Seite 3 - Figure 2 Transducer

Rev. 04/02 -10- D901/M The transducers that are utilized by the D901/M contain piezo electric crystals for transmitting and

Seite 4 - TABLE OF CONTENTS

Rev. 04/02 -11- D901/M Liquid that contains greater than 10,000 PPM [1%] of 30 micron or greater suspended solids or aeratio

Seite 5

Rev. 04/02 -12- D901/M The mounting location and straight pipe requirements for CASE 2 liquid characteristics are the same

Seite 6 - PART 1 - INTRODUCTION

Rev. 04/02 -13- D901/M PART 2 - TRANSDUCER INSTALLATION Before the transducer heads are bonded to the pipe surface, an area

Seite 7

Rev. 04/02 -14- D901/M 2. Wrap the strap around the pipe in the area where the transducers are to be mounted. Leave the st

Seite 8 - PART 1 - SPECIFICATIONS

Rev. 04/02 -15- D901/M the same direction for proper operation. See Figure 7. NOTE: Large pipes may require two people fo

Seite 9

Rev. 04/02 -16- D901/M The enclosure should be mounted in an area that is convenient for servicing, calibration or for obser

Seite 10 - Figure 3

Rev. 04/02 -17- D901/M ♦ Where little vibration exist. ♦ Protected from falling corrosive fluids. ♦ Within ambient tem

Seite 11 - Locations

Rev. 04/02 -18- D901/M enclosure. Secure the transducer cable with the supplied conduit nut. 2. The terminals on the tra

Seite 12 - Figure 4

Rev. 04/02 -1- D901/M This manual contains detailed operating instructions for all aspects of the D901/M instrument. The fo

Seite 13 - Figure 5

Rev. 04/02 -19- D901/M If the 0-1mA output has been installed on the D901/M, the connections are available on the screw ter

Seite 14

Rev. 04/02 -20- D901/M Note: The D901/M flow meter system requires a full pipe of liquid before a successful startup can be

Seite 15

Rev. 04/02 -21- D901/M (covered in the previous sections of this manual) the D901/M can be keypad configured to provide sele

Seite 16 - Figure 7

Rev. 04/02 -22- D901/M The ID key allows the entry of a pipe’s Internal Diameter. The D901/M utilizes the I.D. constant to

Seite 17 - Figure 8

Rev. 04/02 -23- D901/M The FULL SCALE key allows the entry of the highest anticipated fluid velocity. The entry is made as

Seite 18

Rev. 04/02 -24- D901/M The VOL. PULSE key allows the entry of a totalizer exponent. This feature is useful for accommodati

Seite 19 - Transducer Connections

Rev. 04/02 -25- D901/M When utilizing the MGD engineering UNITS, the totalizer defaults to a VOL. PULSE multiplier of 10E3.

Seite 20 - Output

Rev. 04/02 -26- D901/M By applying a CAL value other than 100%, the factory calibrated readings will be altered by the perc

Seite 21 - Instrument Startup

Rev. 04/02 -27- D901/M The DAMP key allows the selection of time duration between display updates. The value selected and e

Seite 22 - Default Configuration

Rev. 04/02 -28- D901/M The RESET key is used for generating a system reset or to reset the accumulated (totalized) flow.

Seite 23 - Pipe I.D. Input

Rev. 04/02 -2- D901/M D. Place each transducer under the mounting strap, 180° apart on the pipe. Ensure that the transduce

Seite 24 - Conversion

Rev. 04/02 -29- D901/M PART 4 - TROUBLE SHOOTING CONDITION POSSIBLE CAUSE Unit does not turn “ON” when power is applied •

Seite 25 - Totalizer Exponent

Rev. 04/02 -30- D901/M PART 4 - TROUBLE SHOOTING FAULT light is ON; low SIGNAL STRENGTH indication (continued) • If GND c

Seite 26

Rev. 04/02 -31- D901/M PART 5 - APPENDICES Appendices Relay Alarm Module Addendum Pipe Dimension Chart: Cast Ir

Seite 27

Addendum — Relay Alarm Module Description: Installation and Operation Manual Addendum Series D901/D90M Alarm Relay Option Origin Date: 7-2

Seite 28 - The TEST Diagnostic

Addendum — Relay Alarm Module Description: Installation and Operation Manual Addendum Series D901/D90M Alarm Relay Option Origin Date: 7-2

Seite 29 - Totalizer

Cement Lining Std./Double ThicknessID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall3 3.96 3.46 0.25 3.40 0.28 3.34 0.31 3.28 0.34 3.22 0.37 3.1

Seite 30 - PART 4 - TROUBLE SHOOTING

Size(Inches)O.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchWallO.D.InchI.D.InchW

Seite 31

ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall ID Wall1 1.315 1.185 0.065 1.097 0.109 1.049 1.049 0.1

Seite 32 - PART 5 - APPENDICES

NominalPipe(Inches)I.D. INCH1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 91 1.05 2.6989 4.0484 5.3978 6.7473 8.097 9.4462 10.796 12.145 13.490 14.8

Seite 33

NominalPipe(Inches)I.D. INCH1 1.5 2 2.5 3 3.5 4 4.5 5 5.5 6 6.5 7 7.5 8 8.5 9FPS TO GPM CROSS - REFERENCE (Schedule 40)18 16.88 697.52 1046.3 1395.0 1

Seite 34

Rev. 04/02 -3- D901/M TABLE OF CONTENTS Pages Quick-Start Operating Instructions 1-2 Introduction General 5 Ap

Seite 35 - Ductile Iron Pipe

Fluid Sound SpeedsOriginal Date: 10/19/99Revision: noneRevision Date: noneFile: I:/dynasonics/dyna_code/tables/doppler ss conversions.xls120.0176921Do

Seite 36 - Cast Iron Pipe

Linseed Oil .925-.939 1770 5803.3 118Methanol (40,41) 0.79 1076 3530.2 72Methyl alcohol (40,44) 0.79 1076 3530.2 72Methylene chloride (3) 1.33 1070 35

Seite 37 - March, 2000

GENERAL TERMS AND CONDITIONS OF SALES 1. PAYMENT – Terms of payment are effective from the actual date of invoice. If, in the Seller’s opinion

Seite 38

Limited Warranty and Disclaimer Dynasonics, div. of Racine Federated Inc. warrants to the end purchaser, for a period of one year from the date

Seite 39

RETURN OF EQUIPMENT/SALES INFORMATION CONTACTS AND PROCEDURES Customer Service/Application Engineer: If you have a question regarding order statu

Seite 40 - Fluid Sound Speeds

D901/M O&M REV 04/02 8635 WASHINGTON AVENUE RACINE, WI 53406 TOLL-FREE IN NORTH AMERICA.: TEL: (800) 535-3569 FAX: (800) 732-8354 TEL: (262) 6

Seite 41

Rev. 04/02 -4- D901/M TABLE OF CONTENTS Pages Instrument Startup and Configuration CAL. Constant Input 26 Displa

Seite 42

Rev. 04/02 -5- D901/M PART 1 - INTRODUCTION The D901/M ultrasonic flow meter is designed to measure the fluid velocity of li

Seite 43

Rev. 04/02 -6- D901/M PART 1 - INTRODUCTION The D901/M employs modular construction and provides electrical safety for the o

Seite 44

Rev. 04/02 -7- D901/M PART 1 - SPECIFICATIONS Ambient Conditions Transmitter - -22 to 160°F [ - 30 to 70 °C], 0-95% relativ

Seite 45 - D901/M O&M REV 04/02

Rev. 04/02 -8- D901/M Outputs 4-20 mA, 600 Ohms max., Isolated. 12 Vdc pulse, 50 mS duration, 0-0.15 Hz to 0-2.5 Hz, user